Remote Operation
Set up the RS-232 Host Port for Remote Control
5
5-15
2.
Connect the selected COM port on the PC to the Calibrator SERIAL 1 FROM HOST
port with a standard null-modem RS-232 cable. (See Appendix C for information on
RS-232 cables and connectors.)
3.
To start the program, open the Test Ports icon from the RS-232 Test group ( see
below).
nn310f.bmp
4.
Verify the Calibrator is turned on and in the reset condition (if in doubt, push
),
then click the Command1 button ( see below).
nn311f.bmp
5.
See that the Calibrator Control Display changes to REMOTE CONTROL (see
below).
nn325f.eps
6.
Click the Command2 button. See that the Calibrator Control Display changes to the
reset condition (see below).
(The Command3 button is used to verify the RS-232 UUT port later in this chapter.)
nn323f.eps
7.
Click on the top-left corner and close the program.