6
6-45
WAVE?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Waveform query) Shows the waveform types for ac outputs. Waveform selections are
SINE (sine wave), TRI (triangle wave), SQUARE (square wave), TRUNCS (truncated
sine wave), or NONE (waveform does not apply).
Responses: <1st waveform> , (SINE, TRI, SQUARE, TRUNCS, NONE)
<2nd waveform> (SINE, TRI, SQUARE, TRUNCS, NONE)
Example:
WAVE?
shows
SQUARE,NONE
Shows
SQUARE
when the ac primary output (Calibrator front panel NORMAL terminals)
is a square wave and
NONE
when there is no secondary output on the front panel AUX
terminals.
ZCOMP
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Impedance Compensation command) Activates or deactivates 2-wire or 4-wire
impedance compensation. For resistance output, compensation is permitted when the
resistance is less than 110 k
Ω
. For capacitance output, compensation is permitted when
the capacitance is equal to or larger than 110 nF. For all other resistances and
capacitances, the compensation is NONE and tries to use other parameters results in the
error message “Can’t change compensation now.” For RTD temperature simulation,
compensation is permitted for all temperatures.
Parameter:
NONE
(Turns off impedance compensation circuitry)
WIRE2
(Turns on the 2-wire impedance compensation circuitry)
WIRE4
(Turns on the 4-wire impedance compensation circuitry)
Example:
ZCOMP WIRE2
Set 2-wire impedance compensation for the Calibrator UUT connection. (Resistance if
the ohms value is less than 110 k
Ω
, capacitance if the farads value is 110 nF or more, or
RTD temperature simulation, any value.)
ZCOMP?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Impedance Compensation query) Shows status of 2-wire or 4-wire impedance
compensation.
Responses:
NONE
(impedance compensation is turns off)
WIRE2
(2-wire impedance compensation is on)
WIRE4
(4-wire impedance compensation is off)
Example:
ZCOMP?
shows
NONE
Shows
NONE
when no impedance compensation is applied to the resistance, capacitance
or RTD output.
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...