SC600 Oscilloscope Calibration Option
Verification Tables
9
9-53
Table 9-16. SC600 Option Marker Generator Verification
Period (s)
Measured Value (s)
Deviation (s)
1-Year Spec. (s)
5 0.0251
s
2 0.00405
s
0.05
3.75E-06
s
0.02
5E-8
0.01
2.5E-8
1e-7
2.5E-13
5e-8
1.25E-13
2e-8
5E-14
1e-8
2.5E-14
5e-9
1.25E-14
2e-9
5E-15
Table 9-17. SC600 Option Pulse Generator Verification: Period
Nominal Value
(V p-p)
Pulse Width
(s)
Period (s)
Measured
Value (s)
Deviation (s)
1-Year Spec.
(s)
2.5 8E-08
2E-06
5E-12
2.5 0.0000005
0.01
2.5E-08
2.5 0.0000005
0.02
5E-08
Table 9-18. SC600 Option Pulse Generator Verification: Pulse Width
Nominal Value
(V p-p)
Pulse Width
(s)
Period (s)
Measured
Value (s)
Deviation (s)
1-Year Spec.
typical (s)
2.5 4.0E-09
2.0E-06
6.2E-9
2.5 4.0E-09
2.0E-05
6.2E-9
2.5 4.0E-09
2.0E-04
6.2E-9
2.5 4.0E-08
2.0E-03
4.4E-8
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...