Table 4-4. Watts Performance, Text Screen
Calibrator Outputs
Performance Limits
Normal
V ac @
60 Hz
Phase
in DEG.
AUX
mV ac @
60 Hz
W/KW VA/KVA
VAR/KVAR
Model 41
Only
Phase
Harmonics
Screen
MIN MAX MIN MAX MIN MAX MIN MAX
5.0 V
0.0
30.0 mV
145
156
145
156
0
4
-2
2
8.0 V
0.0
30.0 mV
234
246
234
246
0
4
-2
2
100.0 V
157.0
150.0 mV
-14.3k -13.3k 14.5k 15.6k 5.4k 6.3k 155 159
100.0 V
157.0
360.0 mV
-37k
-29k
32k
40k
10k
18k
155
159
10.0 V
46.0
1.40 V
9.2
10.2
13.5 14.5 9.6 10.6 44 48
100.0 V
46.0
1.40 V
92
102
135
145
96
106
44
48
1.
Connect the Calibrator to the Model 41 as shown in Figure 4-18.
Note
Voltage is connected to the Model 41 amps input to simulate current clamp
operation (1 mV = 1 A).
5502A
CALIBRATOR
PO
W
ER HARMO
N
ICS
A
N
ALYZER
41
®
gvx028.eps
Figure 4-18. Cable Connections to Verify a 40 Series Watts Function
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...