5502A
Operators Manual
5-44
15
14
13
12
11
10
9
8
0
0
RPTBUSY SETTLED REMOTE
0
UUTBFUL UUTDATA
7
6
5
4
3
2
1
0
HIVOLT
MAGCHG TMPCAL
0
0
0
0
OPER
Set to 1 when a calibration report is being printed to the serial port.
Set to 1 when the output has stabilized to within speclfication or the TC measurement
has settled and is available.
Set to 1 when the 5502A is under remote control.
Set to 1 when data from the UUT port has filled up the UUT buffer.
Set to 1 when there ia data available from the UUT port.
Set to 1 when the 5502A is programmed to a voltage above 33 Volts.
Set to 1 when the output magnitude has changed as a result of another change
(for example RTD_TYPE). This bit is always 0 in the ISR. It changes to 1 only in the
ISCR0 and ISCR1 registers.
Set to 1 when the 5502A is using temporary (non-stored) calibration data.
Set to 1 when the 5502A is in operate, 0 when it is in standby.
RPTBUSY
SETTLED
REMOTE
UUTBFUL
UUTDATA
HIVOLT
MAGCHG
TMPCAL
OPER
gjh049.eps
Figure 5-11. Bit Assignments for the ISR, ISCEs and ISCR
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...