5502A
Operators Manual
4-4
Warm up the Calibrator
When you turn on the Calibrator, let it warm up for 30 minutes to let the internal
components become stable. Specifications given in Chapter 1 apply to a Calibrator that
has completed its warm-up procedure.
If you turn off the Calibrator after warm up and then on again, give it a warm-up period of
no less than two times the length of time it was turned off (maximum of 30 minutes). For
example, if the Calibrator is turned off for 10 minutes and then on again, give it a warm-up
period of no less than 20 minutes.
Use the Softkeys
The five keys to the right of
are called softkeys. Softkey functions come from the
name that shows directly above the key in the Control Display. Push a softkey to change
a value or cause a submenu with new selections to be shown on the Control Display.
Softkey menus are put in different levels, as given in the “Softkey Menu Tree” section in
Chapter 3. You can move back through the menu selections with
. Although
will return you to the top-level menu, it will also reset all volatile settings and return the
Calibrator to 0 V dc in the standby mode. Use
as your main navigation tool to move
through the menu levels.
Use the Setup Menu
Push the front-panel
key to access various operations and changeable parameters.
Most parameters are nonvolatile. This means they will be stored during reset or when you
turn off the power. Chapter 3 shows a map of the menu tree, the parameters, and has a
table of factory-default values.
When you push
from power-up, the display changes to:
nn064f.eps
This is the primary instrument setup menu. The list below tells you about the submenus
that available through each softkey and tells you where you can find more information in
the manuals.
•
CAL
(Calibration)
Opens the calibration menu. You use softkeys in this menu to see the calibration
dates, print a calibration report, calibrate, and to do the Zero calibration routine. Zero
calibration is given in this chapter.
•
SHOW SPECS
(Show Specifications)
Shows published Calibrator specifications for the output value that is currently
selected.
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...