Front-Panel Operation
Sample Applications
4
4-59
c.
Set the Calibrator to 35.0 mA and push
.
d.
Use the output adjustment controls to adjust the Calibrator output for a
measurement of +35.00 mA on the DMM. Make sure that the error shown on the
control display is in specification limits.
e.
Do this step again with 350.0 mA. Make sure the error is in specification limits.
f.
Push the blue key on the DMM to change to ac current measurement.
g.
Set the Calibrator output to 35.0 mA at 60 Hz. Make sure the error is in
specification limits.
h.
Do this step again with the Calibrator set to:
AC Current
Frequency
35.0 mA
1.0 kHz
350.0 mA
60 Hz
350.0 mA
1.0 kHz
i.
Push
on the Calibrator and switch the DMM function switch to
.
j.
Set the Calibrator output to 350
μ
A at 0 Hz and push
. Make sure the error is
in specification limits.
k.
Do this again with 3500
μ
A at 0 Hz.
l.
Push
on the Calibrator and push the blue key on the DMM to switch to ac
measurements.
m.
Change the Calibrator output to 350.0
μ
A at 60 Hz and push
. Make sure the
error is in specification limits.
n.
Do this step again with the Calibrator set to:
AC Current
Frequency
350.0
μ
A 1.0
kHz
3500.0
μ
A 60
Hz
3500.0
μ
A 1.0
kHz
11.
Verify the high current function.
a.
Push
on the Calibrator.
b.
Make sure that the Calibrator is in standby and connect the DMM as shown in
Figure 4-17.
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...