SC600 Oscilloscope Calibration Option
Calibrate the Voltage Amplitude on an Oscilloscope
9
9-11
gjh052.eps
You can push the MODE softkey to move through the functions in the sequence shown,
or you can push
to return directly to the SCOPE menu.
Each menu item is given below:
•
OUTPUT @ SCOPE
Shows the location of the signal output. If the signal does
not show on the oscilloscope, push
. To disconnect the signal, push
.
•
DC <-> AC
Toggles between a dc or ac signal. Push the softkey from the ac signal,
to get the dc equivalent output.
•
1 M
Ω
Toggles the output impedance setting of the Calibrator between 1 M
Ω
and
50
Ω
.
•
TRIG
If you use a square wave to calibrate the external trigger, use this key to
toggle the trigger off and on. When on, the reading will show “/1”, which indicated
that the external trigger is at the same frequency as the volt output. The external
trigger can be useful for many oscilloscopes that have difficulty triggering on low
amplitude signals. You can also toggle the trigger off and on. Push
.
•
V/DIV MENU
Opens the voltage scaling menu, which lets you select the scale of the
signal in volts per division. This menu is given below in detail, under “The V/DIV
Menu.”
•
MODE
Shows you are in Volt mode. Use the softkey to change modes and open the
corresponding menus for the other four oscilloscope calibration modes
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...