SC300 Oscilloscope Calibration Option
Summary of Commands and Queries
10
10-27
SCOPE?
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
Shows the current mode of operation for the oscilloscope. Shows
OFF
if the oscilloscope
is off.
Parameter: (None)
Response: <character> (Shows
OFF
,
VOLT
,
EDGE
,
LEVSINE
,
MARKER,
or
WAVEGEN
.)
TRIG
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
Programs the trigger output BNC of the oscilloscope.
Parameters:
OFF
(Turns off the trigger output.)
DIV1
(Turns on the trigger output. Frequency is the same as the
signal at
SCOPE
output.)
DIV10
(Turns on the trigger output. Frequency is 1/10 of the
signal
at
SCOPE
output.)
DIV100
(Turns on the trigger output. Frequency is 1/100 of the
signal
at
SCOPE
output.)
TRIG?
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
Shows the trigger output setting of the oscilloscope.
Parameters: (None)
Response: <character> (Shows
OFF
,
DIV1
,
DIV10
, or
DIV100
.)
OUT_IMP
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
Programs the output impedance of the oscilloscope.
Parameters:
Z50
(Programs the output impedance of the oscilloscope to
50
Ω
.)
Z1M
(Programs the output impedance of the oscilloscope to
1
M
Ω
.)
OUT_IMP?
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
Shows the output impedance setting of the oscilloscope.
Parameters: (None)
Response: <character> (Shows
Z50
or
Z1M
.)
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...