5502A
Operators Manual
5-8
Verify the IEEE-488 Port
The subsequent procedure verifies IEEE-488 communications between the PC and the
Calibrator with the Win32 Interactive Control utility. This utility is supplied with
National Instruments interface cards for the PC, which are the recommended interfaces.
(See Chapter 8, “Accessories.”) A typical connection is shown in Figure 5-3.
UUT
IEEE-4
88
Port
IEEE-4
88
Port
5502A Cali
b
rator
Controller
IEEE-4
88
Ca
b
le
gjh043.eps
Figure 5-3. IEEE-488 Port Verification
Complete the subsequent procedure to verify IEEE-488 operation with Win32 Interactive
Control.
1.
Complete the “IEEE-488 Port Setup Procedure” in this chapter to set up the
Calibrator for GPIB operation. Note the GPIB Address Port (default is 4).
2.
Connect the PC and Calibrator IEEE-488 ports with a standard IEEE-488 cable. (See
Chapter 8, “Accessories,” for IEEE-488 cables available from Fluke Calibration.)
3.
From the programs menu, select "NI-488.2M software for...(your operating system)".
4.
From the NI488.2M software menu, select "Win32 interactive control".
5.
A DOS window opens with a prompt as shown here:
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...