Remote Operation
RS-232 Interface Overview
5
5-23
Table 5-1. Operation State Transitions
From
To
Front Panel
GPIB Message
Serial Command
Local
Remote
MLA (REN True)
REMOTE
Local with Lockout
LLO
LOCKOUT
Remote
Local
Go to Local Softkey
GTL or REN
False
LOCAL
Remote with
Lockout
LLO
LOCKOUT
Local with Lockout
Local
REN
False
LOCAL
Remote with
Lockout
MLA (REN True)
REMOTE
Remote with
Lockout
Local
REN
False
REMOTE
Local with Lockout
GTL
RS-232 Interface Overview
The two Calibrator RS-232 ports are in accordance with EIA (Electronic Industries
Association) standard RS-232. RS-232 is a serial binary data interchange that operates
from 300 to 9600 baud (selectable), and distances to a maximum of 50 feet. The
Calibrator rear panel SERIAL 1 FROM HOST port is configured as DTE (Data Terminal
Equipment) while the SERIAL 2 TO UUT is configured as DCE (Data Communications
Equipment). See Appendix C for RS-232 cable and connector information. For more
information, see the EIA standard RS-232.
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...