Table 6
Pattern Type
Next Required Action
Sequence
Enable this checkbox if you intend
to download a user-defined
sequence. Patterns generated
within a sequence are defined in
the Sequence Editor.
N O T E
Your instrument also contains user patterns that mimic real data packets and are
designed to stress specific characteristics of your device. For more information, see
3
If you want to select a different pattern for the error detector, clear the checkbox
Error Detector Pattern tracks the Pattern Generator Pattern
. The
Error
Detector Pattern
tab appears. Switch to this tab and repeat step 2.
4
If you have selected "8B/10B Symbol comparison" as an error ratio, then
ensure that the following conditions are met when the checkbox
Error Detector
Pattern tracks the Pattern Generator Pattern
is selected or cleared:
• Condition 1: When pattern tracking check box is selected
– Only user pattern or memory pattern that contains valid 10B symbols
can be loaded in the Pattern Generator.
– Error Detector will follow the Pattern Generator's pattern.
• Condition 2: When pattern tracking check box is cleared
– Any pattern can be loaded in Pattern Generator
– Only user pattern or memory pattern that contains valid 10B symbols
can be loaded in the Error Detector.
5
Click
OK
to finish.
The selected patterns are immediately loaded to the pattern generator and error
detector. The types of patterns that are currently loaded to the pattern generator
and the error detector are always displayed in the status bar at the bottom of the
screen.
3
Setting up Patterns
74
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...