correctly, the VISA Assistant should indicate an instrument named "D4916".
Click
UsbDevice1
or
D4916
.
5
Click the
Formatted I/O
tab.
6
In the
Instr. Lang.
box (Instrument Language), enable
SCPI
.
7
Click the *IDN? button. This generates the identification query in SCPI format.
The instrument should return "Agilent Technologies, N4916A,," and the
software revision.
If you could proceed until here, the N4916A power supply and the processor are
fully functional. If not, the N4916A is probably defective.
Ensure that the User Software can Access the N4916A
1
From the Windows task bar, restore the user interface – click
GUI Agilent 4900
Series
.
2
Press the green
Preset
key close to the front USB connector of the Serial
BERT or from the menu bar select
File
–
Preset Instrument State
.
3
Connect the Data Out of the Serial BERT to the Input of the N4916A.
Solving Problems
9
Agilent J-BERT N4903B High-Performance Serial BERT
495
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...