T I P
With the
Auto Frequencies checkbox, you can adjust the test frequencies
automatically to the selected standard. This is described belowin
.
Setting the Frequency Range
You can freely specify the jitter frequency range and the number of points to be
measured. By default, the test points are suitably distributed within the frequency
range – according to the loaded standard. But you can change the distribution
according to your needs.
1
Switch to the
Frequency
tab.
The measurement's frequency range and the list of test frequencies are
displayed.
If the
Auto Frequencies
checkbox on the
Standard
tab is selected, start and
stop frequency are by default set according to the selected standard.
N O T E
When you change the selection state of the
Auto Frequencies checkbox, press
Apply on the Standard tab to accept the change and to update the frequency list.
8
Jitter Tolerance Tests
456
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...