Preset List
The
Preset
list contains clock rate values that are commonly used by the error
detector.
These presets also modify the loop parameters of the tunable CDR according to
the selected standard. The factory presets consider the parameters as given in the
selected standard.
Click a preset to select it for the error detector clock rate.
Double-click a preset to modify it.
Update with measured data rate
Use the
Update with measured data rate
button to copy the measured data rate
into the Clock Rate field.
Add Preset
Use the
Add Preset
button to add a new preset to the
Preset
list. You can then
enter a value, a unit, and a short description for this new preset.
N O T E
Changes made in this dialog are effective immediately. So, if you press
Cancel, the
settings are still valid.
Edit Preset
Use the
Edit Preset
button to modify the currently selected preset from the list.
You can only edit the presets that are user defined.
Delete Preset
Use the
Delete Preset
button to delete the currently selected preset from the list.
You can only delete presets that are user defined.
Transition Density
The
transition
density is defined by the number of transitions in the incoming data
divided by the total number of bits transmitted. In this field enter the transition
density in (%). Click 'measure' to measure and update the transition density.
This parameter affects the loop bandwidth, and thus must be either entered
correctly or measured. Some standards specify the loop bandwidth for a given
5
Setting up the Error Detector
180
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...