How to View the Results
You can view data from accumulated measurements in several ways:
1
Click the
Accumulated Results
menu item from the
Results
submenu to view
the test results.
For easy comparison, the results of the current and previous accumulation are
“Accumulated Measurements - Reference” on page 378
details on the reported values.
2
Examine the measurement log files on an external PC.
This allows to access all data for detailed processing and analysis on a PC.
Also, you can track the status of all indicators (such as DATA LOSS and CLK
LOSS) during the accumulation period. If you want to view results from an
accumulation period prior to the current and previous test, you can view the
respective log file, too.
How to Analyze the Results
Accumulated measurements can help you determine the cause of bit errors. Bit
errors can be caused by a variety of problems. In addition, instrument settings can
affect how errors are measured and displayed. The sync mode setting has the
greatest effect.
• Analyze the basic accumulated results and log files. Possible observations that
may occur are:
– Errors became constant (and remained constant for a period of time) or
number of errored 0's was greater than errored 1's, see
(More Errored 0's than 1's)” on page 473
– Sync loss seconds were measured, see
Accumulated Measurements - Reference
The
Accumulated Results
window presents data from the current and previous
accumulation period.
N O T E
Data from the current and all previous accumulation periods is available in the
measurement log files, if correctly set up.
This window contains a graph and five tables:
7
Evaluating Results
378
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...