threshold is set below or above the data eye, the output at Aux Out will be
constant high or low, respectively.
4
Click
OK
to finish the
Trigger and Aux Out
setup.
Trigger and Aux Output - Reference
The available options in the
ED Trigger & Aux Output
window are:
Trigger
You have the following options to configure the signal at the Trigger Out port:
•
Clock Divided by n
The trigger is set every
n
bits.
Clock
Trigger Out
•
Pattern
The trigger signal is sent every time when the complete pattern has been
received. This refers to the single pattern repetitions, not the 512-bit blocks.
•
Filler Detected
The usage of the enhanced error detector trigger is based on the 8B/10B data
coding. The filler removal function is configured with up to four different detect
words which can be 1, 2, or 4 symbols and also allows for wildcards. This
flexibility is used to load detect words that is typically required for DUT
handshakes. The filler removal unit triggers upon a detected word which is
used with the pattern generator AUX input to advance a pattern sequence and
thus implements a simple handshake. The enhanced error detector trigger
capability is only available when the 8B/10B coding support is activated. Only
the trigger pulse timing is compatible with the pattern generator Aux input for
sequence advance. It takes 2048 bit time period to process the trigger and the
signal is active only for half of this period. Hence next trigger will be processed
only after this period.
N O T E
The
Filler Detected option will be available only if you have selected the error ratio
based on 8B/10B symbol comparison.
•
SKPOS Detected
5
Setting up the Error Detector
204
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...