next. The chosen method impacts the measurement duration. The specified
parameters influence both measurement duration and precision.
You can choose one of six search methods:
• Downwards linear
• Downwards logarithmic
• Upwards linear
• Upwards logarithmic
• Binary search
• Extended binary
For a description of these methods refer to
“Vertical Search Methods” on page
.
For linear search, you can set the step size.
•
Downwards linear
starts from the maximum (which depends on the jitter
frequency). A step size of 100 mUI, for example, may result in a sequence of
1000 UI, 999.9 UI, 999.8 UI, and so on.
The test for one frequency stops when the BER limit is met or zero amplitude
is reached.
•
Upwards linear
starts from the minimum. A step size of 100 mUI, for example,
will result in a sequence of 0.1 UI, 0.2 UI, 0.3 UI, and so on.
The test for one frequency stops when the BER limit is exceeded or the
maximum amplitude is reached.
For logarithmic search, you can set the minimum amplitude and the ratio. The ratio
is the factor by which the present amplitude value is multiplied to calculate the
next amplitude.
•
Downwards logarithmic
starts from the maximum (which depends on the jitter
frequency). A ratio of 0.666, for example, may result in a sequence of 1000 UI,
666 UI, 444 UI, and so on.
The test for one frequency stops when the BER limit is met or the specified
minimum amplitude is reached.
•
Upwards logarithmic
starts from the specified minimum amplitude. A ratio of
1.5, for example, may result in a sequence of 0.1 UI, 0.15 UI, 0.225 UI, 0.3375
UI, and so on.
The test for one frequency stops when the BER limit is crossed or the maximum
amplitude (which depends on the jitter frequency) is reached.
For binary search, you can set the accuracy.
Jitter Tolerance Tests
8
Agilent J-BERT N4903B High-Performance Serial BERT
445
Linear search
Logarithmic search
Binary search
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...