Periodic Jitter 1 Parameters
The Periodic Jitter is characterized by:
• Amplitude
• Frequency
• Waveform
A graph indicates whether the chosen setting of frequency and amplitude is
tolerable.
The maximum peak-to-peak
Amplitude
is limited by the free capacity of the chosen
delay line.
The maximum
Frequency
of the periodic jitter source depends on the chosen
waveform. Refer to the technical specifications.
You can switch between sine, rectangular, or triangular jitter source waveforms.
Periodic Jitter 2 Parameters
The Periodic Jitter 2 has the following modes:
• Periodic Jitter 2 – Constant Jitter
• Periodic Jitter 2 – Constant Amplitude Sweep
• Periodic Jitter 2 – Variable Amplitude Sweep
Jitter Tolerance Tests
8
Agilent J-BERT N4903B High-Performance Serial BERT
421
Amplitude
Frequency
Waveform
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...