Exporting Measurement Results
To export the measurement results into a .txt file for later use in external
spreadsheet applications:
1
After your measurement has finished and the results are displayed, select
Export Data
from the
Analysis
menu.
2
In the
Save as
dialog box, select the path and file name for the .txt file.
3
Press
Save
to export the data to the specified destination.
Loading Measurements
To load previously saved measurements:
1
Select
Open Measurement
from the
File > Open
menu.
2
Use the file dialog box to locate the desired .jcp file and press
Open
to load the
measurement.
Accessing Comment and Condition Text
To access the comment or condition text of a measurement:
• Double-click
Modify Comment
below the graphical display.
This opens the
Jitter Measurement Comment
dialog box.
You can enter a comment. The condition is automatically added by the system.
You can also read the full comment or condition in case only part of the text is
displayed in the result table.
Jitter Tolerance Compliance - Reference
Jitter Tolerance Compliance determines whether a device under test maintains a
desired BER at defined jitter levels. The jitter levels can be derived from predefined
standards.
Standard Specification
The drop-down list contains all predefined standards, including user-defined
standards. The shape of the selected standard is shown in the preview area. The
8
Jitter Tolerance Tests
462
Agilent J-BERT N4903B High-Performance Serial BERT
Standard Selection
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...