Eye Diagram Screen
This graph gives the picture of the Eye Diagram Screen.
Spectral Jitter
Spectral Jitter - Concepts
The Spectral Jitter measurement allows you to analyze the jitter inherent in the
output signals of your device under test (DUT) as a function of the frequency. This
measurement can be used for investigation of the behavior of the DUT, for example
to identify crosstalk effects.
Spectral Jitter Analysis
The Spectral Jitter measurement provides a method for investigating the spectrum
of the jitter, i.e. its power distribution over frequency.
Deterministic jitter can be caused by repetitive data patterns. It can also be caused
by internal or external periodic effects.
6
Advanced Analysis
342
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...