Note that a de-emphasis value of 6 dB causes a 50 percent amplitude reduction.
Check the Output of the Pattern Generator with a Scope
If the N4916A does not produce a signal, check the Data Out of the Serial BERT.
• Connect the Data Output of the pattern generator to an oscilloscope.
N O T E
As long as the N4916A is in "connected" state (after pressing the
Connect button),
the pattern generator's output is programmed to a fixed level. The amplitude is set
to 1.2 V, the offset to 0. This yields a signal of ±0.6 V.
The oscilloscope should show the following picture:
If you see only a straight line, ensure that the pattern generator is not disabled and
that it is set up to generate a pattern (e.g. PRBS).
Check the Output of the Pattern Generator with the Error Detector
If you have connected the N4916A to a Serial BERT you can of course check the
output of the pattern generator by means of the error detector.
1
Set up the pattern generator as described above: Set the Data Offset to 0 V and
the Amplitude to 1 V.
2
Set up the error detector as described above.
Solving Problems
9
Agilent J-BERT N4903B High-Performance Serial BERT
499
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...