Table 31
S.No
Parameter
Description
10
Jitter RMS
A standard deviation of the
crossing point histogram
with respect to BER
threshold.
Note: The Jitter RMS is
calculated according to
the BER Threshold set in
the BERT from Calculate
Measurement Parameters
of the View Tab.
11
Cross Voltage
Crossing percentage is a
measure of the amplitude
of the crossing points
relative to the 1 level and 0
level.
Note: The Cross Voltage is
calculated according to
the BER Threshold set in
the BERT from Calculate
Measurement Parameters
of the View Tab.
12
Signal to Noise Ratio
In signal-to-noise, the
"signal" is the information
power of the signal
indicated by the difference
between the ‘1’, and ‘0’
level. The "noise" is the
combined standard
deviations of the ‘1’ level
spread, and the ‘0’ level
spread.
13
Extiction Ratio
Extinction ratio is the ratio
of the 1 level power and the
0 level power of an eye
diagram.
Note: The Extinction Ratio
is calculated using the
Dark Level and Conversion
Gain of the View Tab.
6
Advanced Analysis
340
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...