Error
This indicator turns red when errors are detected. View the
BER
bar or BER Results
to see the nature of the errors.
BERT connected looped back or to your device:
• Stable errors caused by the error add function: Turn error add off.
• Stable errors caused by false sync: Select a lower sync threshold BER.
• Variable and high errors may be caused by faulty connectors/cables.
BERT connected to your device:
• Stable or variable errors can also be caused by your device. Correct the
problems with your device.
• If the problems are not clear, then accumulate measurement data and analyze
results.
Measurement Problems
Measurement Problems - Concepts
Accumulated measurements can help you determine the cause of bit errors. Bit
errors can be caused by a variety of problems. In addition, instrument settings can
affect how errors are measured and displayed. The sync mode setting has the
greatest effect.
9
Solving Problems
472
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...