Instantaneous Measurements - Reference
This section describes all elements of the
BER Results
window and
8B/10B
Comparison Results
windows and their various status indicators:
BER Results Window
This window displays the actual (
BER
) or cumulative (
AccumBER
) BER results:
• The
BER
is the current BER, calculated upon a period of 200ms.
• The
AccumBER
shows either the accumulated BER of the current
accumulation, or - if no accumulation is running - the results of the most recent
accumulation.
This enables you to monitor real-time BER behavior as you do things such as to
manually adjust the sampling point, to add errors, or to make adjustments to your
device.
Click this button to toggle between
BER
or
AccumBER
results.
This area displays the actual error count or cumulative error count.
N O T E
The
BER bar below the main menu bar also displays the actual BER. You can drag
the small yellow alarm threshold mark to change the BER Alarm Threshold.
7
Evaluating Results
366
Agilent J-BERT N4903B High-Performance Serial BERT
Accumulated/Actual Button
Error Count/Accum Error Count
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...