The
Sample Point
in the eye diagram moves horizontally as you change the
value.
T I P
While adjusting the data input delay, you can monitor the
BER bar on the analyzer
display, or listen to the BER warning tones. Higher BERs may indicate that you are
measuring at the edges of the data eye.
N O T E
Pattern synchronization can be lost during sampling point adjustment. If
Auto
Sync is enabled, the Serial BERT will resynchronize the patterns automatically. If
the manual synchronization mode is selected, you need to synchronize the error
detector by clicking the
Pattern Sync icon and then the Sync Now button.
How to Adjust the 0/1 Decision Threshold
1
Click
Sampling Point Setup
menu item from the
ED Setup
submenu.
2
Clear the
Avg. 0/1 Threshold
checkbox.
3
Click
0/1 Threshold Center
to move the sampling point vertically towards the
center of the eye at the selected delay. You may click
Cancel
at any time.
4
Now you can use the numeric keypad or front panel knob to manually adjust
the value in the
0/1 Threshold
field.
The
Sample Point
in the eye diagram moves vertically as you change the value.
T I P
While adjusting the 0/1 threshold, you can monitor the
BER bar on the analyzer
display, or listen to the BER warning tones. Higher BERs may indicate that you are
measuring at the edges of the data eye.
N O T E
Pattern synchronization can be lost during sampling point adjustment. If
Auto
Sync is enabled, the Serial BERT will resynchronize the patterns automatically. If
the manual synchronization mode is selected, you need to synchronize the error
detector by clicking the
Pattern Sync icon and then the Sync Now button.
How to Enable the Bit Recovery Mode
The bit recovery mode (BRM) allows you to measure the relative BER without
knowing the incoming data pattern.
To enable the bit recovery mode:
1
Click
Sampling Point Setup
menu item from the
ED Setup
submenu.
Setting up the Error Detector
5
Agilent J-BERT N4903B High-Performance Serial BERT
197
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...