data rate / Acquisition Depth
For example: If you have a data rate of 2.5 GHz and an
Acquisition Depth
of
128 Kbit, the frequency resolution is 19.0735 kHz.
The relations are illustrated in the following figure:
N O T E
A high
Acquisition Depth requires a high degree of computational effort and hence
time.
•
Sample Point Offset
By default, the sampling point for the measurement is positioned 0.5 clock
periods or UI ahead of the present analyzer sampling point.
If the sampling point is set in the middle of the eye, this corresponds to the
transition point of the incoming signal. If you have set the sampling delay
manually, this is the time offset from that point.
This option allows you to fine-tune the sampling point. The unit is UI (unit
intervals). This makes the setting independent of the present clock frequency.
Set the FFT window for the
FFT Calculation
. The following windows are provided:
• Uniform
• Hanning (sometimes also called Hann)
• Hamming
• Blackman
The "uniform" window is no window at all. This setting supplies the original error
record to the FFT.
6
Advanced Analysis
354
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...