The maximum peak-to-peak
Amplitude
is limited by the chosen frequency.
The maximum
Frequency
of the rSSC jitter source depends on the amplitude.
Sinusoidal Jitter Parameters
Sinusoidal jitter (SJ) is independent from periodic jitter. It is generated by
modulating the clock that is used for data generation. It can therefore not be
enabled, if SSC or rSSC is active.
Sinusoidal jitter is characterized by
• Amplitude (peak-to-peak)
• Frequency
There is a dependency between frequency and acceptable amplitude. A graph
indicates whether the chosen setting of frequency and amplitude is acceptable.
The maximum peak-to-peak
Amplitude
is limited by the chosen jitter frequency.
Note that high-frequency sinusoidal data jitter can be generated by the
Periodic
Jitter
function or by applying an
External
jitter source.
8
Jitter Tolerance Tests
420
Agilent J-BERT N4903B High-Performance Serial BERT
Amplitude(p-p)
Frequency
Amplitude
Frequency
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...