3
Terminate PG Data /Out with 50Ohm
4
Terminate PG Clock /Out with 50Ohm
5
Ensure that the error detector follows the generated pattern.
– Press the
Auto Align
button.
– The error detector should finish the auto alignment successfully and report
a BER of 0.000.
– Manually adjust the ED sample delay by 166 ps.
– Press
Sync Now
.
– The error detector should report a BER of 0.000.
Check the Output of the Pattern Generator with a Scope
If the N4876A does not produce a signal, check the Data Out of the Serial BERT.
• Connect the Data Output of the pattern generator to an oscilloscope.
N O T E
As long as the N4876A is in "connected" state (after pressing the
Enable button),
the pattern generator's output is programmed to a fixed level. The amplitude is set
to 0.8 V, the offset to -0.4 V.
The oscilloscope should show the following picture for Data Out as well as for Aux
Data Out:
9
Solving Problems
516
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...