Period
The accumulation period cannot be specified for
Manual
tests.
•
Time
Select this option to configure the error detector to accumulate bit errors for a
specific period of time.
You can enter the desired time period in the
Days
,
Hr
(Hours),
Mn
(Minutes),
and
Sec
(Seconds) fields. You may use the numeric keypad or front panel knob.
The minimum value is 1 second and the maximum value is 99 days, 23 hours,
59 minutes, and 59 seconds. When the selected time has elapsed, the
accumulation ends.
•
Number of Errors
Select this option to configure the error detector to accumulate bit errors until
a specified number of errors has been measured.
The number of errors can be set to
10
,
100
, or
1000
. When the selected number
of errors has occurred, the accumulation ends after the next full second.
•
Number of Bits
Select this option to configure the error detector to accumulate bit errors until
the specified number of bits have been examined.
The number of bits can be set from
1E7
to
1E15
. When the selected number of
bits have been exceeded, the accumulation ends.
N O T E
The actual test period may contain more bits than specified.
N O T E
The accumulation period cannot be changed while accumulation is progress.
Browse
Click this button to browse for the directory where you wish to save the
measurement log files.
N O T E
Changes to the parameters of the accumulation setup do not take effect until the
start of the next accumulation.
5
Setting up the Error Detector
222
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...