Then the oscilloscope should show an output eye like the following:
If you see a display like above everything is ok.
If you see merely one fixed level, either the N4916A or the output of the Serial
BERT is defective.
Check the Output of the N4916A Using the Error Detector
If you have connected the N4916A to a Serial BERT and have no oscilloscope at
hand, you can use the error detector for checking the output of the N4916A.
1
Set up the pattern generator as described above: Set the Data Offset to 0 V and
the Amplitude to 1 V. Set the De-Emphasis to 6 dB.
2
Press
ED Setup
and open the
Sampling Point Setup
page.
3
Set up the error detector:
– Press the
Edit
button to enter the
ED Input Setup Dialog
.
– Set the
Input
to "Normal".
– Set the
Input Range
to –1 V to +1 V.
– Set the
Threshold
to "Averaging".
– Set the
Termination
to 0 V.
Solving Problems
9
Agilent J-BERT N4903B High-Performance Serial BERT
497
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...