In this example, we expect the signal voltages to be between -1.75V and -0.95V.
The
Sample Threshold
values proposed above cover this range well.
The
Resolution
is the distance between the measurement points when the
threshold moves from the low to the high level. A resolution of 10mV results
in 100 measured points per Volt.
Note that we have disabled the
Edge Resolution Optimization
.
4
Press
OK
to close the
Properties
dialog box.
5
Press the
Start
button to execute the measurement.
The measurement is run, and the result window shows the bit error rates
measured at 100 threshold levels from -1.85V up to -850mV.
Advanced Analysis
6
Agilent J-BERT N4903B High-Performance Serial BERT
267