Table 31
S.No
Parameter
Description
7
Eye Width
The horizontal
measurement of the eye
opening at a specified BER
Threshold.
Note: The Eye Width is
calculated according to
the BER Threshold set in
the View Tab. Eye Width is
calculated as per the
selection of either Width
at Crossing Point or
Custom defined Width.
8
Eye Height
Measures the vertical
opening of an eye diagram
with respect to BER
threshold. This determines
“eye closure” due to noise.
Note: The Eye Height is
calculated according to
the BER Threshold set in
the View Tab. Eye Height is
calculated with 'Height at'
in the View tab.
9
Jitter P-P
Full width of the histogram
at the eye crossing point
with respect to BER
threshold.
Note: The Jitter P-P is
calculated according to
the BER Threshold set in
the BERT from Calculate
Measurement Parameters
of the View Tab.
Advanced Analysis
6
Agilent J-BERT N4903B High-Performance Serial BERT
339
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...