Select the Delay Line
To select the delay line for (sinusoidal) periodic jitter 1 and 2 and BUJ, press
610ps
button. The LED will turn green if the 610ps delay line is selected.
Note that sRJ can not be enabled if the 610ps delay line is selected.
N O T E
The 610ps delay line cannot be used for bit rates above 3.37 GHz.
Specify the Jitter Components
In this section, an example of periodic jitter 1 is shown to specify the jitter
components.
To specify the jitter components:
1
Activate a jitter source (press the corresponding button).
2
Set the most commonly used parameter (typically
Amplitude
and
Frequency
)
directly.
3
Press
Edit
button to access all parameters of the corresponding jitter type.
This opens a window which allows you to set all parameters available for the
given jitter type.
Jitter Tolerance Tests
8
Agilent J-BERT N4903B High-Performance Serial BERT
409
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...