Error Ratio - Reference
8B/10B Presets Dialog
The
8B/10B Presets
dialog allows you to enter a short description, values for Filler
Primitives definition and Alignment Symbol for the new preset. This dialog contains
the following elements:
Filler Primitives Definition
Filler Primitives are inserted or deleted for clock tolerance compensation. These
are not compared and therefore cannot be counted as errors. Filler Primitives
contain filler symbols. A maximum of 4 alternative filler primitives can be used.
Each filler primitive can consist of upto 4 filler symbols. Filler symbols are separated
by a comma (,).
Example: For SAS we have 4 filler primitives each consisting of 4 filler symbols.
K28.5,D10.2,D10.2,D27.3
K28.5,D7.0,D7.0,D7.0
K28.5,D1.3,D1.3,D1.3
K28.5,D27.3,D27.3,D27.3
The following are the supported standards along with their filler symbols:
5
Setting up the Error Detector
190
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...