frequency range and the jitter amplitude range of the standard is given in the
preview. A description is displayed at the bottom of the
Standard
tab.
The last entry in the list is
Browse…
. Select this entry to load a file that contains
“Setting the Standard” on page 454
To aggravate test conditions, the standard's test jitter amplitude can be increased
by
Margin
percent. The display shows a black line for the standard and a green
line for the new jitter amplitudes.
Select the
Auto Frequencies
checkbox to set frequencies that are recommended
for the selected standard. This list contains, for example, the edge frequencies of
the standard.
T I P
You can change single frequency points. Press
Apply to accept the selected Auto
Frequencies setting and switch to the Frequency tab. In Manual mode, you can
edit the automatically calculated frequencies. If you do so, the
Auto Frequencies
checkbox is automatically deselected.
Enter the number of test frequencies (
Number of Steps
) either on this tab, or on
the
Frequency
tab.
This is the creator's description of the selected standard. Standards provided by
Agilent are described by the following items:
T I P
Use the scroll bar if not all information is displayed.
•
Document
that describes the standard in detail. The document name usually
indicates the date or revision of the standard.
• Recommended settings for the test pattern like
Bit Rate
range,
Pattern
types
and voltage level ranges:
– Voltage
Level from Standard
: This is the peak-to-peak differential input
level range for the DUT as given in the standard.
– Voltage
Level at PG (recom)
: This is the recommended output level at the
pattern generator to achieve the minimum voltage
Level from Standard
.
N O T E
This recommendation depends on the simulated intersymbol interference (ISI): The
minimum voltage
Level from Standard is only met if the trace to simulate the ISI is
selected as in the description item
Additional Jitter. Other settings of the ISI can
change the DUT's input level up to four times.
Jitter Tolerance Tests
8
Agilent J-BERT N4903B High-Performance Serial BERT
463
Margin
Auto Frequencies / Number of Steps
Description
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...