N O T E
The
BER Threshold will usually be 0, since you want the measurement to fail if a
single bit was received in error.
The pass/fail threshold applies to all measurement points. An
icon indicates all measurements where the bit error rate is higher than this
threshold, as shown in the figure below:
View Tab
The following parameters can be set on the
View
tab of the
Properties
dialog box:
You can analyze for:
•
All Errors
To display all errors.
•
Errors if 0s Expected
To display the errors if "0" is expected, but "1" received.
•
Errors if 1s Expected
To display the errors if "1" is expected, but "0" received.
You can set the
Timing Unit
for the display of the relative time of the measurement
points on the Parameters tab. Choose between
Unit Interval
and
Seconds
to select
the unit.
6
Advanced Analysis
314
Agilent J-BERT N4903B High-Performance Serial BERT
Analyze
Grid
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...