
N O T E
During measurement logging, the Serial BERT logs data in ten-second intervals.
Your log file may be missing up to the last ten seconds of data. To avoid this
condition, accumulate for 10 seconds longer than desired.
Accumulated Measurements - Procedures
This section describes how to run accumulated measurements and how to analyze
the results:
How to Run Accumulated Measurements
To run accumulated measurements:
1
Ensure that the instrument is set up correctly and that the error detector is
already in synchronization.
2
Click the
Accumulation Setup
menu item from the
ED Setup
submenu to check
the settings made for the data accumulation.
“Error Accumulation - Procedures” on page 220
3
To begin accumulation, press the
Start Accum
button on the front panel.
This causes the following events:
• If
Prompt for File Name
has been selected during setup, you will be asked to
enter a name for the log file before accumulation begins.
• The data display in the
Accumulated Results
window is reset.
• The recording of data to the measurement log file is reset.
How to Stop Accumulated Measurements
If you specified a single or repeat test, the test runs until the specified end condition
is reached. If you selected a manual test, it will run until you stop it manually.
• Press the
Stop Accum
button on the front panel to stop the test immediately.
N O T E
Errors are only counted when the signal at the Gate In port is low.
Evaluating Results
7
Agilent J-BERT N4903B High-Performance Serial BERT
377
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...