During a jitter tolerance compliance test, jitter with the specified jitter amplitude
is sequentially applied at a number of frequency points in the frequency range of
interest.
The receiver checks for transmission errors and measures the bit error rate.
The graphical display shows the results.
A green circle indicates that the BER at this point did not exceed the target BER.
A red cross indicates that the BER at this point exceeded the target BER. If this
happens, the test is marked as “failed”.
A table below the graph provides details about the measurement conditions like
date/time, data pattern, bit rate, and added jitter components.
The measurement results for each point can also be exported as a numerical list
in text format.
The maximum jitter amplitude the Serial BERT can generate is also frequency
dependent, as illustrated below.
8
Jitter Tolerance Tests
448
Agilent J-BERT N4903B High-Performance Serial BERT
Measurement Results
Instrument Capability
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...