In
Manual
mode, you can edit the automatically calculated frequencies. You can
also enter your own list of frequencies, with a minimum of two frequencies
specified. For example, you can jump between low and high jitter frequencies.
Enter the frequencies in Hz, separated by commas, for example:
2e4, 1.75e6, 6.334e5, ...
T I P
Press
Apply if you want to keep the setting of frequencies for the current test run,
but if you still want to change the standard or BER specification. If all properties are
set as desired, press
OK.
BER Specification
The
Target Bit Error Ratio
defines the pass/fail criteria of the jitter tolerance
compliance test. For every test point along the standard curve, the test is failed if
the measured BER lies above the
Target Bit Error Ratio
.
The
Verification Method
defines under which circumstances the measurement
proceeds from one frequency to the next. You can choose between two
Verification
Methods
:
Setting a
Confidence Level
is generally recommended. This ensures that neither
too few nor too many bits are captured. See also
.
If you enable
Absolute Bits/Errors
, the measurement proceeds to the next jitter
frequency as soon as either the number of received bits or the number of errors is
reached.
N O T E
Ensure that you capture a number of bits that suffices to verify (with the desired
probability) that a measured point has a BER below the
Target Bit Error Ratio.
The relax time allows the DUT to recover in case it has lost synchronization due
to an excessive high bit error rate. The next BER measurement starts after this
time has elapsed.
To run the Jitter measurements you need to make sure that Global Jitter, Sinusoidal
Jitter and Periodic Jitter are enabled.
Jitter Tolerance Tests
8
Agilent J-BERT N4903B High-Performance Serial BERT
465
Manual mode
Target Bit Error Ratio
Confidence Level
Absolute Bits/Errors
Relax Time
Optimal Measurement Setup
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...