Error Accumulation
Error Accumulation - Concepts
The Serial BERT instantly starts measuring the BER when the error detector
receives data. However, to perform tests that can be repeated and compared, you
can collect the measurement data in several ways and save it to a log file.
The available options are to test:
• for a specified time,
• until a certain number of errors occurred,
• until a specified number of bits have been sent,
• until you stop it manually.
Error Accumulation - Procedures
To set up the Serial BERT for error accumulation:
1
Click the
Accumulation Setup
menu item from the
ED Setup
submenu
2
Select the desired
Activation Mode
for the type of test you want to run
(
Manual
,
Single
, or
Repeat
test).
3
If you selected a
Single
or
Repeat
test, you need to specify the accumulation
Period
(
Time
,
Number of Errors
, or
Number of Bits
).
4
Use the
Measurement Logs
area to specify whether you want save the results
in a log file.
5
The log files will be saved in the
Log Directory
displayed at the bottom of the
dialog box. Use the
Browse
button to select a different directory.
6
Click
OK
to apply your selections and close the
Accumulation Setup
dialog box.
For more information on how to run error accumulation tests, see
Accumulated Measurements” on page 377
5
Setting up the Error Detector
220
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...