Choose between a binary or hexadecimal display of the pattern. The current
setting is shown in the status bar below the pattern window.
Note that in hexadecimal view the captured error can only be located as being
one of a group of four bits. To display which of the four bits is the errored bit,
you have to switch to the binary display.
•
Exp. / Cap.
(Expected/Captured) Toggle the data view between the display of the expected
data, i.e. the data you defined on the error detector, and the captured data, i.e.
the data that was captured at the error detector. The current state is displayed
in the title bar of the pattern grid.
•
Run Select
Toggle the data view between the current run and the previous run.
After a measurement run is successfully finished, the display is automatically
switched to the current run. The current state is displayed in the title bar of the
pattern grid.
•
First Error
Jumps to the first errored bit in the pattern.
•
Prev. Error
Jumps to the previous errored bit in the pattern.
•
Next Error
Jumps to the next errored bit in the pattern.
•
Last Error
Jumps to the last errored bit in the pattern.
Explanation of the Results
The results are shown in various ways:
The graphic view highlights the located bit errors. The currently selected bit is
highlighted in red, all other errored bits are written in red.
The status bar below the graphic view provides the following information:
•
Status
: The status indicates if the measurement is running or has ended. If the
measurement has ended, it indicates if the measurement ended successfully
or was aborted.
•
Position
: The position of the current errored bit is shown.
•
Hex/Bin
: This field indicates how the data is shown (bin or hex format).
6
Advanced Analysis
306
Agilent J-BERT N4903B High-Performance Serial BERT
Graphic View
Status Bar
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...