7
Set up the error detector to match the input range and the termination with the
pattern generator’s levels:
– Select an Input Range from –2 V to 0 V
– Set the Data Termination to –2 V
– Set the Alignment BER Threshold to 1E–6
– Set the Clock Setup to Clock Data Recovery to get the error detector’s clock
from the incoming data stream.
8
Enable the pattern generator outputs by pressing the
0V (Disable)
button.
How to Initiate an Eye Diagram Measurement?
To run the eye diagram measurement:
1
Click the
Eye Diagram
menu item from the
Analysis
submenu.
2
Click the
Start
button to execute the measurement.
The measurement runs, and the eye diagram along with the Automated Eye
Parameter Measurement get updated constantly.
The following graph shows the eye diagram.
3
To stop the measurement click on Abort.
Advanced Analysis
6
Agilent J-BERT N4903B High-Performance Serial BERT
321
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...