Then the oscilloscope should show an output eye like the following:
If you see a display like above everything is ok.
If you see merely one fixed level, either the N4916B or the output of the Serial
BERT is defective.
Check the Output of the N4916B (De-Emphasis Part) Using the Error
Detector
If you have connected the N4916B to a Serial BERT and have no oscilloscope at
hand, you can use the error detector for checking the output of the N4916B.
1
Connect the ED Data-In to the Data Out of N4916B
2
Connect the ED Clock In to the PG Clock Out
3
Terminate PG Data Out with 50Ohm
4
Terminate PG Clock Out with 50Ohm
5
Ensure that the error detector follows the generated pattern.
– Press the
Auto Align
button.
– From the
Analysis
panel select the
Eye Diagram
page.
– Press the
Start
button.
Solving Problems
9
Agilent J-BERT N4903B High-Performance Serial BERT
505
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...