The following figure clarifies the behavior in the range between 615Mbit/s and
620 Mbit/s:
615
Bitrate [Mbit/s]
620
Limitations do not apply
in this range
Limitations apply
in this range
For the error detector the following rules apply:
• For low frequencies, you cannot use the automatic data alignment functions
(
Auto Align
and
Data Center
). Instead you need to align the error detector
manually. For instructions, please refer to the Online Help, go to the
Error
Detector Setup
section, select
Sampling Point Setup
, and switch to the
Procedures
information.
• There are restrictions to the available sampling point delay values.
The error detector can vary the sampling point delay only within a range of 0ps
to 1.6129ns (relative to the clock signal). For frequencies above 620 Mbit/s,
this range is sufficient to cover the complete clock cycle (= 1 unit interval).
For lower frequencies, the maximum sampling point delay is smaller than the
clock cycle. Therefore, the sampling point cannot be set everywhere within the
clock cycle.
Clock 500 Mbit/s
Valid Range for
Sampling Point
2 ns
Valid Sampling Point
0
1.6129 ns
If you cannot find the optimum sampling point in the valid range of sampling
points, you can switch from the rising to the falling clock edge. For this purpose
activate the
Clock Falling Edge
checkbox on the
Sampling Point Setup
screen
of the error detector setup.
Clock 500 Mbit/s
Rising Edge
Sampling Point Delay
Falling Edge
Sampling Point Delay
2 ns
Covered Range
Setting up the Error Detector
5
Agilent J-BERT N4903B High-Performance Serial BERT
177
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...