You can enable the input after reducing the voltage to a safe level.
Delay Control Input - Procedures
N O T E
The following procedure does not apply to Serial BERT on which the calibrated and
integrated jitter injection option J10 is installed. For such instruments refer to
“Jitter Setup - Procedures” on page 407
.
To set up a jitter tolerance test on an instrument without option J10:
1
Press
PG Data Setup
in the
PG Setup
menu.
2
Disable the Delay Control In port by clearing the
Delay Control Input
checkbox.
3
Connect the device that supplies the jitter signal to the Delay Control In port.
4
Enable the port again by selecting the
Delay Control Input
checkbox.
T I P
While monitoring the BER you can use this checkbox to switch jitter on and off and
view the effect on the BER.
Delay Control Input - Reference
This checkbox in the
PG Data Setup
window is used to disable or enable the Delay
Control In port, where you can supply an external signal to add jitter to the output
signal.
4
Setting up the Pattern Generator
156
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...