• No other advanced measurement (Output Timing, Output Levels, etc.) can be
running.
• Error Location Capture can only run when the BER Location Mode is set to
more than one bit (for example, all bits, or a block with a length > 1).
Error Location Capture - Procedures
This section shows how to set up and perform an Error Location Capture
measurement. As an example first add a couple of errors to an alternating pattern
and then capture their position.
This requires the following steps:
• Preparing the measurement (see
“How to Prepare the Error Location Capture
)
• Demonstrating the measurement (see
“How to Demonstrate the Error Location
Capture Measurement” on page 305
How to Prepare the Error Location Capture Measurement
To prepare an Error Location Capture measurement to detect the location of an
errored bit in an alternating pattern:
1
Disable the pattern generator outputs by pressing the
0V (Disable)
button in
the
PG Setup -> Data Output
screen.
2
Use a shielded cable to connect the pattern generator's Data Out port and the
error detector's Data In port.
3
Use a second shielded cable to connect the pattern generator's Clock Out port
and the error detector's Clock In port.
4
Terminate all non-connected pattern generator output ports with 50 Ω.
5
Make sure to preset the instrument state before running the measurement:
Select
Preset Instrument State
from the
File
menu.
6
Enable the pattern generator outputs by pressing the
0V (Disable)
button.
7
Press
Sync Now
and then
Auto Align
to find the optimum sampling point. Check
that the synchronization and alignment were successful. None of the error
indicators should show red.
The resulting BER should be zero.
6
Advanced Analysis
304
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...