N O T E
Ensure that the timing adjustment has to be run at least once when either the
N4903B, N4876A or the cables connecting both are being exchanged in the test
setup.
Also, ensure that the timing adjustment has to be repeated whenever the operating
temperature differs by more than 5
o
Celsius from the temperature at the previous
timing adjustment.
How to Enable/Disable M8061A Multiplexer with De-emphasis
Function
To enable/disable the M8061A Multiplexer with De-emphasis function:
1
Select the
Mux with DeEmphasis
function from the given list. It opens the
M8061A Connection
dialog.
The following image shows the an example of M8061A.
2
Click
Enable
button. It enables multiplexer function that is connected between
Data and Aux Data Output of J-BERT's Pattern Generator and the input of the
M8061A.
Setting up External Instrument(s)
2
Agilent J-BERT N4903B High-Performance Serial BERT
49
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...