However, if you select error ratio based on "8B/10B Symbol Comparison",
follow the below steps.
5
Click
cBER Setup
.
It opens a
cBER Setup
dialog. This dialog offers the choice to calculate cBER
based on either SER, FER, FSR, ISR, or DER with either automatic conversion
factor or manually specified conversion factor.
C A U T I O N
When cBER is calculated from FSR, DER, ISR, in this case the cBER might be lower
than the sync threshold, even if the cBER based on SER or FER is higher than sync
threshold. This affects the automatic re-sync behavior and a manual "Sync Now"
is required.
6
Select the predefined standards listed in the
Setup
window.
N O T E
In order to achieve Symbol Lock, the pattern loaded in PG must have the same
Comma Symbol as selected from the standard.
7
Click
Add Preset
button to open
8B/10B Presets
dialog. It allows you to add a
new preset to the
Setup
list. You can then enter a short description, values for
filler symbols and align symbols for the new preset. For more details, see
Ratio - Reference” on page 190
.
N O T E
Changes made in this dialog are effective immediately. So, if you press
Cancel, the
settings are still valid.
8
Select a preset in the
Setup
list and click
Delete Preset
button to remove it
from the list. You can only delete presets which are defined by the user.
Setting up the Error Detector
5
Agilent J-BERT N4903B High-Performance Serial BERT
187
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...