– Press the
Auto Align
key on the front panel
– Switch to the
Accumulated Results
Screen
– Press the
Start Accum
key on the front panel to start the BER
measurement
– For each measurement write down the number of errors at the end of the
measurement
– The correct Input Timing is selected by the following rule:
Look for the BER measurement that produced no error, and that has the
biggest distance to a measurement with errors and select the Input Timing
setting that was used in this measurement.
Example: Input Timing Setting 2 has the biggest ‘distance’ to a setting that
produced errors, no matter if checking in ‘forward’ or ‘backward’ direction
through the table. So this is the setting that has to be used.
9
Solving Problems
482
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...